Skip to content

Public FDA record

Z-1437-2026: Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A;

Recall record Z-1437-2026 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1437-2026
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Feb 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.

Source facts

Action
On January 15, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected products in your inventory. 3. Olympus requests that you acknowledge receipt of this letter through the recall web portal provided. 4. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your WA22810A outer sheath. 5. If you have further distributed this product, please forward this notification to other users.
Initiated
2024-09-13
Root cause
Under Investigation by firm
Distribution
Distribution US nationwide.
Recall number
Z-1437-2026
Recall status
Open, Classified
CDRH record ID
218207
FDA classified
2026-02-20
Recall event number
98300

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.