Open original FDA record ↗Recall record
- FDA identifier
- Z-1438-2012
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Apr 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin were evaluated and measured against print specifications.
Source facts
- Action
- Biomet sent an "URGENT MEDICAL DEVICE SAFETY ALERT" letter dated March 20, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to continue monitoring implanted patients for disassociation of the pin component and report to Biomet. The Alert included a complete listing of device invoice history and patient information (if supplied). Contact the firm for confirmation of receipt of the notice and for questions at 574-372-3983.
- Initiated
- 2012-03-20
- Root cause
- Component design/selection
- Terminated
- 2013-02-07
- Distribution
- Distributed in California and Pennsylvania.
- Recall number
- Z-1438-2012
- Recall status
- Terminated
- CDRH record ID
- 107270
- FDA classified
- 2012-04-19
- Recall event number
- 61089