Open original FDA record ↗Recall record
- FDA identifier
- Z-1438-2021
- Company in FDA record
- Heartware, Inc.
- FDA record creation date
- Apr 20, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603
Source facts
- Action
- Beginning 26-Feb-2021, an Urgent Medical Device Recall notification was disseminated to consignees who Medtronic records indicate have received affected product, UPS 2-day delivery. The notices provided updated information for the instructions for use and patient manual.
- Initiated
- 2021-02-26
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Italy, Great Britain, Germany, Japan.
- Recall number
- Z-1438-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186391
- FDA record created
- 2021-04-20
- Recall event number
- 87604