Improper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
Source facts
Action
The recalling firm has decided to inspect all of the installed units for this condition and where necessary replace the missing clip. The recalling firm has issued an Update Instructions AX011/09/S dated 03.09, which provides instructions on how to perform a visual inspection of the Mavig Portegra 2 arm for the presence of the safety clip and the installation instructions on how to properly install the clip if it is missing.
Initiated
2009-05-01
Root cause
Other
Terminated
2009-09-01
Distribution
Nationwide Distribution.
Recall number
Z-1439-2009
Recall status
Terminated
CDRH record ID
81722
FDA classified
2009-06-16
Recall event number
51947