Skip to content

Public FDA record

Z-1439-2014: Philips HeartStart XL+ Defibrillator/Monitor Model number: 861290, automatic external defibrillator.

Recall record Z-1439-2014 for Philips Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1439-2014
Company in FDA record
Philips Medical Systems, Inc.
FDA classification date
Apr 9, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A component on the HeartStart XL+ Therapy Board could malfunction potentially affecting the ability to deliver therapy

Source facts

Action
Philips Healthcare issued an Urgent-Medical Device Correction letter dated March 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This Field Safety Notice is intended to inform you about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer/user in order to prevent risks for patients or users " the actions planned by Philips to correct the problem. A component on the HeartStart XL+ Therapy Board could malfunction potentially affecting the ability to deliver therapy. Please see the attached Field Safety Notice that provides information on how to identify affected devices and instructions on actions to be taken. Please follow the ACTION TO BE TAKEN BY CUSTOMER / USER section of the notice. To correct this issue Philips will provide a replacement therapy board to customers with affected units free of charge. This voluntary correction has been reported to the appropriate regulatory agencies. If you need any further information or support concerning this issue, please contact your local Philips representative at Philips representative contact details to be completed by the KM / country. Philips apologizes for any inconveniences caused by this problem. Further questions please call (978) 687-1501.
Initiated
2014-03-31
Root cause
Reprocessing Controls
Terminated
2015-10-21
Distribution
Worldwide Distribution - US Distribution including the states of IN and WA., and the countries of : CHINA FRANCE GERMANY ITALY MALAYSIA PANAMA SAUDI ARABIA TURKEY and UNITED KINGDOM.
Recall number
Z-1439-2014
Recall status
Terminated
CDRH record ID
126503
FDA classified
2014-04-09
Recall event number
67857

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗