Reason stated in the source

Incorrect Reagent Status-When a new cartridge of Infinity Lithium Reagent is loaded on to the UniCel DxC System, the Reagent Status screen will incorrectly show 'Days Left' as 21. The correct Reagent Status for a new Lithium reagent cartridge is 14 days. Lithium reagent used past 14 days may produce low quality control and/or patient results.

Source facts

Action
A Product Correction Action (PCA) letter was sent the week of May 15, 2006 to all Unicel DxC customers informing them not to leave any Lithium reagent cartridge onboard a UniCel DxC System past 14 days. They also were instructed to complete and return the enclosed response form. The Letter was sent by US Mail.
Initiated
2006-05-15
Root cause
Other
Terminated
2013-04-08
Distribution
Nationwide and Canada
Recall number
Z-1440-06
Recall status
Terminated
CDRH record ID
46712
FDA classified
2006-08-26
Recall event number
35646

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗