Skip to content

Public FDA record

Z-1440-2009: Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging

Recall record Z-1440-2009 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1440-2009
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jun 16, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Improper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.

Source facts

Action
The recalling firm has decided to inspect all of the installed units for this condition and where necessary replace the missing clip. The recalling firm has issued an Update Instructions AX011/09/S dated 03.09, which provides instructions on how to perform a visual inspection of the Mavig Portegra 2 arm for the presence of the safety clip and the installation instructions on how to properly install the clip if it is missing.
Initiated
2009-05-01
Root cause
Other
Terminated
2009-09-01
Distribution
Nationwide Distribution.
Recall number
Z-1440-2009
Recall status
Terminated
CDRH record ID
81723
FDA classified
2009-06-16
Recall event number
51947

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗