Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.
Source facts
Action
Carl Zeiss Meditec, Inc sent a Customer Notification letter to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that firm's sales representative will be scheduling an on-site visit, informing them of the affected product and providing instructions on the recall. For question call 925-580-5371.
Initiated
2013-04-18
Root cause
Under Investigation by firm
Terminated
2013-06-26
Distribution
US Nationwide Distribution.
Recall number
Z-1440-2013
Recall status
Terminated
CDRH record ID
117898
FDA classified
2013-05-31
Recall event number
65109