Skip to content

Public FDA record

Z-1440-2020: Monitor for the Functional Imaging System for Magnetic Resonance Imaging System

Recall record Z-1440-2020 for Invivo Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1440-2020
Company in FDA record
Invivo Corporation
FDA record creation date
Mar 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Insufficient shielding of Philips SensaVue DVI Adapter for computer monitor.

Source facts

Action
On February 5, 2020 Invivo sent out URGENT Field Safety Notice identifying problem and under what circumstances it can occur; the actions that should be taken by the customer / user in order to prevent risks for patients or users and the actions planned by Philips to correct the problem. PRODUCTS Products manufactured between O8DEC2O1S and 09MAY2019 are in scope of this field correction and can be identified by the manufacturing date printed in the SensaVue Cart label. ACTION TO BE TAKEN BY CUSTOMER/ USER: Although the level of emissions is undetectable without specialized equipment and pose no risk to user or patient safety, customers shall contact their service provider. Customers can request Part Number 459801750971 PACKAGE. FCD SensaVue DVI Adapter to obtain a new, compliant DVI adapters. Please follow the instructions in the kit to replace two (2) adapters in the back of the operator computer. After replacement, complete and e-mail the Action Notification Report included in the kit and discard the old adapters. ACTIONS PLANNED BY PHILIPS A replacement DVI adapter has been selected, tested and passed the requirements for emissions contained within IEC 60601-1-2 3rd/4th edition. This new DVI adapter will be used to replace non-compliant DVI adapters in the field shipped from 06DEC2018 to 09MAY2019. FURTHER INFORMATION AND SUPPORT If you need any further information or support concerning this issue, please contact your local Philips representative: North America: 1-877-Invivo-1 option 3
Initiated
2020-02-06
Root cause
Component change control
Terminated
2022-05-04
Distribution
U.S Distribution: AK, AL, CA, CO, FL, IL, IN, KY, MA, MI, NY, OH, OR, PA, TN, TX and WV. *** Foreign: Australia, Canada, France, United Kingdom, Italy, Japan, Netherlands, Saudi Arabia, India and Thailand.
Recall number
Z-1440-2020
Recall status
Terminated
CDRH record ID
179798
FDA record created
2020-03-04
Recall event number
84957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.