Open original FDA record ↗Recall record
- FDA identifier
- Z-1441-2014
- Company in FDA record
- Remington Medical Inc.
- FDA classification date
- Apr 9, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).
Source facts
- Action
- Remington Medical Inc sent an Field Safety Notice dated March 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please be aware that the typographical error described above is included in the FL-601-97 Instructions for Use (RM-0154 Rev 05) provided with the lots that you have received as identified below. RMI has corrected the typographical error in the revised FL-601-97 Instructions for Use (RM-0154 Rev 5.1) which is attached to this communication. Please incorporate the information contained in the attachment with the existing FL-601-97 labeling (Instructions for Use) provided with the product. If you have distributed these products further, please forward copies of this notice and the revised Instructions for Use to any recipients as appropriate. If you have any questions regarding this communication, please contact RMIs sales representatives or customer service team at 800-989-0057, Monday - Friday, between the hours of 8 am - 4pm EST. Patient safety is the highest priority for Remington Medical, Inc. We appreciate your assistance with this matter and regret any inconvenience this may cause.
- Initiated
- 2014-03-05
- Root cause
- Labeling Change Control
- Terminated
- 2018-04-09
- Distribution
- USA (nationwide) Distribution including Canada
- Recall number
- Z-1441-2014
- Recall status
- Terminated
- CDRH record ID
- 126209
- FDA classified
- 2014-04-09
- Recall event number
- 67752