Open original FDA record ↗Recall record
- FDA identifier
- Z-1441-2020
- Company in FDA record
- Advanced Bionics, LLC
- FDA record creation date
- Mar 4, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.
Source facts
- Action
- On 4/4/2019, the recalling firm sent a letter to its customers via FedEx asking that the customer return the corrupted USB drive to the firm.
- Initiated
- 2019-04-04
- Root cause
- Software in the Use Environment
- Terminated
- 2022-06-27
- Distribution
- US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI
- Recall number
- Z-1441-2020
- Recall status
- Terminated
- CDRH record ID
- 179431
- FDA record created
- 2020-03-04
- Recall event number
- 84800