Skip to content

Public FDA record

Z-1442-2022: NanoSight NS300

Recall record Z-1442-2022 for MALVERN PANALYTICAL LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1442-2022
Company in FDA record
MALVERN PANALYTICAL LTD
FDA record creation date
Aug 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product found to not comply to Class 1 Laser safety requirement.

Source facts

Action
Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and manual addenda with revised safety instructions.
Initiated
2022-04-07
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide - Worldwide Distribution
Recall number
Z-1442-2022
Recall status
Open, Classified
CDRH record ID
194993
FDA record created
2022-08-01
Recall event number
90597

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗