When using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.
Source facts
Action
The firm issued product notification letter February 20, 2008, to its consignees. The notification alerts customers to the behavior, requires them to post the notification and notify users not to use the Freehand ROI feature until they are upgraded to a corrected software version. Philips will deliver a software upgrade to correct the problem.
Initiated
2008-01-29
Root cause
Software design
Terminated
2008-09-25
Distribution
Worldwide Distribution --- USA including states of MN, CO, and PA, and country of Germany.
Recall number
Z-1443-2008
Recall status
Terminated
CDRH record ID
69032
FDA classified
2008-08-16
Recall event number
46942