Open original FDA record ↗Recall record
- FDA identifier
- Z-1444-2022
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA record creation date
- Jul 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.
Source facts
- Action
- An Urgent Medical Device Removal Notification issued September 08, 2020 sent via USPS certified mail. The notice identifies the affected part numbers S4C CROSS CONNECTOR SW112T, SW113T, SW114T, SW115T, SW116T and SW117T and asks distributors and customers to identify existing inventory, quarantine, and return affected inventory. If further distributed, identify and notify customers. The notice includes an acknowledgement form to be completed and returned to Aesculap Inc. via email at qa-recalls@aesculapusa.com. Questions are to be directed to Lindsay Chromiak, Manager, QA Compliance and Vigilance (610) 984-9072.
- Initiated
- 2020-09-08
- Root cause
- Device Design
- Terminated
- 2024-04-10
- Distribution
- United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.
- Recall number
- Z-1444-2022
- Recall status
- Terminated
- CDRH record ID
- 194651
- FDA record created
- 2022-07-21
- Recall event number
- 90502