Open original FDA record ↗Recall record
- FDA identifier
- Z-1445-2014
- Company in FDA record
- Alcon LenSx, Inc.
- FDA classification date
- Apr 9, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Alcon LenSx received reports of unexpected downward motion of the gantry.
Source facts
- Action
- Alcon LenSx Inc advised customers of a Medical Device Correction being initiated for specific serial numbers of the LenSx Laser System for use in cataract surgery for the creation of corneal cuts/incisions, anterior capsulotomy and laser phacofragmentation. Alcon sent customer notification letters via mail during the week of March 17, 2014. The letter titled "Medical Device Correction", advised customers of the correction and provided: the product's description with serial number(s), description of the potential condition, advise on action to be taken by the user, transmission of the notice, and a contact reference person. The letter also contained a section for customers to sign and return the page of the Notice to Alcon.
- Initiated
- 2014-03-14
- Root cause
- Device Design
- Terminated
- 2014-10-16
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1445-2014
- Recall status
- Terminated
- CDRH record ID
- 126213
- FDA classified
- 2014-04-09
- Recall event number
- 67755