Skip to content

Public FDA record

Z-1446-2020: Covidien Force TriVerse Electrosurgical Devices, Product numbers FT3000 and FT3000DB, UDI 10884524000084

Recall record Z-1446-2020 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1446-2020
Company in FDA record
Covidien Llc
FDA classification date
Feb 25, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sterile packaging may potentially be compromised, and use of products may result in increased risk for infection.

Source facts

Action
On February 10, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed of the packaging issue and instructed to do the following: 1. Please immediately quarantine and discontinue use of the affected item codes and lots listed in the notice. 2. Please return affected product as indicated in the notice. All unused products from the affected item codes and lots must be returned. 3. If you have distributed Covidien Force TriVerse electrosurgical devices listed on Attachment A, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Initiated
2020-02-10
Root cause
Process control
Terminated
2024-04-26
Distribution
Nationwide domestic distribution. Foreign distribution to Australia, Canada, Austria Belgium Canary Islands Czech Republic Denmark Finland France Germany Greece Hungary Italy Jordan Kazakhstan Kenya Lithuania Luxembourg North Macedonia Poland Portugal Reunion Romania Russian Federation Serbia Slovakia Spain Sweden Switzerland United Kingdom No U.S. distribution of Kits. Foreign distribution in EMEA only. (Europe, Middle East and Africa.)
Recall number
Z-1446-2020
Recall status
Terminated
CDRH record ID
179848
FDA classified
2020-02-25
Recall event number
84991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗