Skip to content

Public FDA record

Z-1448-06: ADVIA 1650 Chemistry System, Automated Clinical Chemistry Analyzer - human serum, plasma and urine tests.

Recall record Z-1448-06 for Bayer Healthcare, LLC (Diagnostics Division). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1448-06
Company in FDA record
Bayer Healthcare, LLC (Diagnostics Division)
FDA classification date
Aug 29, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Low QC recovery observed on carbamazepine (CARB) results immediately following a Gentamicin (GENT) assay or a Digoxin (DIG) assay. When GENT or DIG precedes CARB, the CARB test result is artificially low (as much as -25%). PHNY was determined to also exhibit a similar carryover effect on CARB. All others were found not to produce the same reagent probe-based carryover effect.

Source facts

Action
Bayer HealthCare LLC issued a Support Bulletin to all affected Bayer HealthCare LLC Branches on/about 2/17/06 in order to inform them of the issue and to provide them with instructions for this ''field correction''. Customer Bulletins were also issued to the Branches both in the United States and elsewhere for communication with the affected customers.
Initiated
2006-02-17
Root cause
Other
Terminated
2012-05-24
Distribution
Worldwide including USA and Puerto Rico.
Recall number
Z-1448-06
Recall status
Terminated
CDRH record ID
47080
FDA classified
2006-08-29
Recall event number
35880

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗