Open original FDA record ↗Recall record
- FDA identifier
- Z-1448-2014
- Company in FDA record
- Pega Medical Inc.
- FDA classification date
- Apr 10, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Male Components in this lot are made of material with lower strength than manufacturing specifications.
Source facts
- Action
- Pega Medical sent an Urgent Field Safety Notice-Recall letter dated February 7, 2014, to their customers via email. The letter identified the affected product, the problem, and the action to be taken by the customer. Customers were instructed to identify the affected devicee, verify if it was used in a patient and place the affected device in quarantine immediately if not used. If not used, return the affected device to Pega Medical Inc. to the address provided. Pega's customer notification stated "If used, it is recommended to continue the normal patient follow up. Customers with questions should call (450) 688-5144 ext. 242 or email egarcia@pegamedical.com
- Initiated
- 2014-02-07
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-04-23
- Distribution
- Worldwide Distribution - USA Nationwide in the states of Missouri, Texas, and the country of Germany.
- Recall number
- Z-1448-2014
- Recall status
- Terminated
- CDRH record ID
- 126028
- FDA classified
- 2014-04-10
- Recall event number
- 67675