Reason stated in the source
The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution was added.
Source facts
- Action
- FertiPro sent an Urgent: Medical Device Recall letter dated December 22, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to check if they have stock of the concerned lots and provide a confirmation that they have destroyed the affected products. Distributors were asked to inform their customers and instruct them to do the same. For further questions you may call (858) 824-0888.
- Initiated
- 2014-12-22
- Root cause
- Component design/selection
- Terminated
- 2015-05-05
- Distribution
- US Distribution to the state of : OR., and Internationally to Argentina.
- Recall number
- Z-1448-2015
- Recall status
- Terminated
- CDRH record ID
- 132286
- FDA classified
- 2015-04-15
- Recall event number
- 70131