When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.
Source facts
Action
Urgent Notification Software Update Notification letters dated June 2022 were sent to customers.
Stryker Instruments will be updating the software on all SurgiCount+ Tablets (0794-001-010) distributed on or before May 1, 2022.
Actions to be Taken
1. Immediately review this Software Update Notification. Advise OR staff, and all other personnel who may use this product, of the information in the Notification.
2. A local Stryker representative will schedule and perform the software update on each SurgiCount+ Tablet. A form, documenting the software update, will be completed by the Stryker Sales Representative, and a copy will be provided to your facility.
3. Please complete the enclosed Business Reply Form to confirm receipt of this notification. Return to Jordan.Greegor@stryker.com.
We apologize for any inconvenience this action may cause your facility. Please forward a copy of this letter to any other personnel within your facility that you deem appropriate.
Initiated
2022-06-15
Root cause
Under Investigation by firm
Terminated
2024-04-05
Distribution
The products were distributed to the following US states: KS and ME.
Recall number
Z-1452-2022
Recall status
Terminated
CDRH record ID
194574
FDA record created
2022-07-22
Recall event number
90487