Skip to content

Public FDA record

Z-1453-2022: Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2

Recall record Z-1453-2022 for NuVasive Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1453-2022
Company in FDA record
NuVasive Inc
FDA record creation date
Jul 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.

Source facts

Action
On 06/20/22, Field Safety notices were delivered in person, emailed, or mailed to hospital and distributor consignees who were informed of the following: 1) Complete and return the consignee confirmation form. 2) Return of affected devices is not required. However, if you would like these items exchanged, please contact your firm representative to request a Return Material Authorization number (RMA). 3) This recall notice should be shared with anyone who needs to be aware within your organization or shared with any organization where the potentially affected products have been transferred. Customers with questions, complaints, or who want to report adverse effects can contact the firm at complaints@nuvasive.com
Initiated
2022-06-20
Root cause
Labeling Change Control
Distribution
US: CA, WA, TX, IN, MO, MI, FL, VA, OR. OUS: DE, NZ, SG, UK
Recall number
Z-1453-2022
Recall status
Open, Classified
CDRH record ID
194600
FDA record created
2022-07-22
Recall event number
90500

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗