Z-1455-2017: XTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatme

Recall record Z-1455-2017 for Strata Skin Sciences, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1455-2017
Company in FDA record
Strata Skin Sciences, Inc.
FDA classification date
Mar 27, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.

Source facts

Action
STRATA Skin Sciences, Inc., planned action to bring defect into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 1000 domestic units and 612 international units will consist of the following: 1. Send Field Correction notification to all consignees by January 30, 2017, 2. Verify all consignees received the notification by February 10, 2017, 3. Ship all Rear Panel Boards to consignees by May 1, 2017, and 4. All corrective actions will be performed at no cost to the purchasers For further questions call Customer service phone no.: (877) 449-8722
Initiated
2017-01-23
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-06-05
Distribution
USA (nationwide) Distribution
Recall number
Z-1455-2017
Recall status
Terminated
CDRH record ID
153927
FDA classified
2017-03-27
Recall event number
76697

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗