Open original FDA record ↗Recall record
- FDA identifier
- Z-1455-2017
- Company in FDA record
- Strata Skin Sciences, Inc.
- FDA classification date
- Mar 27, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.
Source facts
- Action
- STRATA Skin Sciences, Inc., planned action to bring defect into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 1000 domestic units and 612 international units will consist of the following: 1. Send Field Correction notification to all consignees by January 30, 2017, 2. Verify all consignees received the notification by February 10, 2017, 3. Ship all Rear Panel Boards to consignees by May 1, 2017, and 4. All corrective actions will be performed at no cost to the purchasers For further questions call Customer service phone no.: (877) 449-8722
- Initiated
- 2017-01-23
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-06-05
- Distribution
- USA (nationwide) Distribution
- Recall number
- Z-1455-2017
- Recall status
- Terminated
- CDRH record ID
- 153927
- FDA classified
- 2017-03-27
- Recall event number
- 76697