Skip to content

Public FDA record

Z-1455-2025: Brand Names: MyDay Toric, LensCrafters One Day Premium Daily Disposable Contact Lenses For Astigmatism, P

Recall record Z-1455-2025 for CooperVision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1455-2025
Company in FDA record
CooperVision, Inc.
FDA classification date
Mar 26, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

a limited number of lots were manufactured with an incorrect cylinder power.

Source facts

Action
On February 27, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to direct consignees. We request you take the following actions: 1. Please examine your inventory/product shelf for any of the provided affected lot numbers. The lot number can be found on the back panel of the carton below the expiration date. (Carton lots 19346502595028, 19346502613028, 21545502813013, and 21512501331524) 2. While trials lenses are not impacted by this action, if your patients have discarded the carton, the affected product can be identified by examining the lot number printed on the blister. This number is different from the carton lot number. (Blister lots 6461501582, 6461501590, 6448502235, 6472500679) The lot number can be found on the front of the blister label below the expiration date. Below is an example and where to locate the lot number on both a carton and blister. Stop any further distribution of product from the designated lot numbers; 3. We recommend authorized distributors contact your accounts and Eye Care Practitioners contact your patients who have received the affected product with the enclosed draft Consumer Communication Letter, to be completed by the Eye Care Practitioner s office. 4. Please instruct any patients to return any affected product to their Eye Care Practitioner for replacement or credit. 5. Within the next five (5) business days, please complete the attached Correction Response Form and fax it to 888-385-3217 or scan and email to: info@coopervision.com. If you have no lenses from these specified lots, please put 0 in Quantity ; 6. Dispose of the affected product in your inventory or that is returned to you. 7. Your account will be credited once the form is received. 8. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting Program, either online, by regular mail, or by fax. 9. If your patients have questions or complaints, please have them contact our cu
Initiated
2025-02-27
Root cause
Process control
Distribution
Domestic: CA, CO, FL, IL, KY, MI, NC, NE, NJ, NY, OH, OR, UT, VA, WA. International: Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Ireland, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Spain, Sweden, Switzerland, and United Kingdom.
Recall number
Z-1455-2025
Recall status
Open, Classified
CDRH record ID
212775
FDA classified
2025-03-26
Recall event number
96396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗