Open original FDA record ↗Recall record
- FDA identifier
- Z-1455-2026
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Feb 20, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Complaints of the ceramic tip of the resection sheath breaking have been received.
Source facts
- Action
- An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at Cynthia.Ow@Olympus.com or by phone at 947-999-3203.
- Initiated
- 2025-11-11
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide Distribution.
- Recall number
- Z-1455-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218173
- FDA classified
- 2026-02-20
- Recall event number
- 98288