Z-1457-2024: Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Recall record Z-1457-2024 for Covidien, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1457-2024
Company in FDA record
Covidien, LP
FDA classification date
Apr 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Source facts

Action
On February 14, 2024, the firm notified domestic customers and notified international customers starting February 19, 2024 of the issue via URGENT: MEDICAL DEVICE RECALL letters. Actions to be taken by the customer: 1. Identify and quarantine all unused impacted product listed above. 2. See Attachment A for guidance on identifying affected lots. 3. Return all quarantined devices to Medtronic. Please contact rs.covidienfeedbackcustomerservice@medtronic.com for the Return Goods Authorization (RGA). 4. Credit for the returned affected devices will be issued based on the RGA number. 5. If purchased from a distributor, contact your distributor directly to arrange for the return of the devices back to your distributor. 6. Please complete and return the enclosed Customer Confirmation Form to rs.gmbmitgfca@medtronic.com even if you do not have unused inventory. 7. Pass on this notice to all those who need to be aware within your organization or to any organization where the listed affected lots may have been transferred or distributed. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) and Medtronic Technical Services at 800-255-6774, option 2 to provide information regarding those events. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Field Representative or Customer Service at 800-962-9888 Option 2.
Initiated
2024-02-14
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
Recall number
Z-1457-2024
Recall status
Open, Classified
CDRH record ID
206390
FDA classified
2024-04-01
Recall event number
94128

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗