Microbiological media for antibiotic susceptibility testing gives false test results.
Source facts
Action
The recalling firm notified distributors and direct accounts by fax, e-mail, and letters on 07/28/2005. The notifications advised of false resistant results with control organisms known to be oxcacillin susceptible. Customers were instructed to discontinue use of the lot and discard for replacement. Distributors were requested to discontinue distribution and provide consignee list for subsequent notification. Response forms are included with the recall letters.
Initiated
2005-07-28
Root cause
Other
Terminated
2005-10-17
Distribution
Product was sold to distributors and direct account end-users worldwide including government facilities.
Recall number
Z-1458-05
Recall status
Terminated
CDRH record ID
41123
FDA classified
2005-08-31
Recall event number
32943