Open original FDA record ↗Recall record
- FDA identifier
- Z-1458-2014
- Company in FDA record
- Biomet 3i, LLC
- FDA classification date
- Apr 15, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to residual machining fluid identified by discoloration (darker in appearance) on the external dental implant surface.
Source facts
- Action
- Biomet 3i sent an Urgent Medical Device Notice dated September 9, 2013, to all affected customers. The notice identified the product, problem and actions to be taken. Each customer was instructed to check their respective inventory for the affected product and to return any unused quantities along with the attached Recall Response Form. Biomet 3i will issue replacement product at no charge. For questions or concerns, call Biomet 3i at 1-800-342-5454.
- Initiated
- 2013-08-29
- Root cause
- Process control
- Terminated
- 2014-10-17
- Distribution
- Worldwide Distribution - USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Korea (South), Lebanon, Liechtenstein, Lithuania, Mexico, Netherlands, Poland, Portugal, Paraguay, Saudi Arabia, Sweden, Singapore, Turkey, Taiwan, Uruguay, Vietnam and South Africa.
- Recall number
- Z-1458-2014
- Recall status
- Terminated
- CDRH record ID
- 125833
- FDA classified
- 2014-04-15
- Recall event number
- 67609