Dose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration of dose to a patient.
Source facts
Action
Consignees were notified by a Medical Device Notification letter sent on 01/29/2008. The letter advised users to cease use of the device until a corrected version of the software is available. A representative will contact users to upgrade the application once the revised software has been validated and released for distribution.
Initiated
2008-01-30
Root cause
Software design
Terminated
2010-08-18
Distribution
Nationwide Distribution including states of NC, FL and CA.
Recall number
Z-1459-2008
Recall status
Terminated
CDRH record ID
68342
FDA classified
2008-08-02
Recall event number
46852