Open original FDA record ↗Recall record
- FDA identifier
- Z-1460-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 29, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Calibration solution for ABL700 Blood Gas Analyzers is labeled with an incorrect bar code. The bar code identified on the Cal Solution 2 product is actually the bar code for the rinse solution.
Source facts
- Action
- Representatives of the recalling firm were informed of this problem and were asked to visit the firm''s customers and retrieve any of the suspect product found in inventory at each customer location for customer credit. The representatives were instructed to complete this action via a ''Management Cover e-mail issued on 10/31/2005. According to the Management Cover Memo, these actions were scheduled for completion by firm representatives on or before 12/15/2005.
- Initiated
- 2005-10-31
- Root cause
- Other
- Terminated
- 2012-04-17
- Distribution
- The recalled product was distributed to customers in following states nationwide: AL, AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV and WY
- Recall number
- Z-1460-06
- Recall status
- Terminated
- CDRH record ID
- 46573
- FDA classified
- 2006-08-29
- Recall event number
- 35663