Incorrect Image Orientation: The image orientation tag may not be set correctly if the patient has been scanned with an orientation other than Head First Spine. When such images are imported to a treatment planning system, they may appear with either a mirrored or a rotated view.
Source facts
Action
Consignees were notified by an Urgent Medical Device Correction letter issued on 02/05/2008. The letter provided recommended actions for users to follow in order to avoid the software error. The letter informed users that a service pack is being created to correct the issue and that the recalling firm will notify users when it becomes available. For additional information, contact 888-827-4265.
Initiated
2008-01-03
Root cause
Software design
Terminated
2009-09-30
Distribution
Worldwide Distribution--USA including states of CA, MA and PA, and countries of UK, Chile, Belgium, Canada and Germany.
Recall number
Z-1460-2008
Recall status
Terminated
CDRH record ID
68508
FDA classified
2008-07-30
Recall event number
46919