Skip to content

Public FDA record

Z-1460-2012: ELITE CARE CLINER 6940(STD)/6950(X-LG) w/ Swing Arm products are labeled in part: "****ELITE CARE CLINER

Recall record Z-1460-2012 for Winco Mfg., LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1460-2012
Company in FDA record
Winco Mfg., LLC
FDA classification date
Apr 25, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Winco Manufacturing LLC is recalling the Care Cliner, XL Care Cliner, Drop Arm Care Cliner, Elite Care Cliner, and XL Elite Care Cliner chairs with nylon casters distributed between 04/01/2009 and 08/31/2009. Certain 5" nylon casters shipped with the chairs may be defective. There is a potential of the caster's axle working itself out of the assembly under unconventional conditions. Should this

Source facts

Action
Winco Mfg. sent an "URGENT DEVICE RECALL" letter dated February 1, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were asked to complete the Caster Recall Replacement Form and return it to the firm's Customer Service Department via fax to 352-854-9544. Contact the Customer Service Department at 800 237-3377 for questions regarding this notice.
Initiated
2012-02-06
Root cause
Device Design
Terminated
2012-07-27
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada and Singapore.
Recall number
Z-1460-2012
Recall status
Terminated
CDRH record ID
108146
FDA classified
2012-04-25
Recall event number
61457

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗