Open original FDA record ↗Recall record
- FDA identifier
- Z-1460-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential issue and possible hazard to patients when using the AXIOM Vertix MD Trauma systems. In rare cases, steel ropes inside the lift column of the system can be defective without triggering the safety lock, which can result in the U-arm dropping down unexpectedly during movement in vertical direction, potentially causing serious injury.
Source facts
- Action
- Siemens sent a Safety Advisory Notice dated March 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised as a first check it is strongly recommended for the users to check whether metallic dust or rubbed off parts of metal are visible underneath the lifting column or around the system. If this is the case, it is strongly recommended to immediately stop using the Vertix MD Trauma system and call the local Siemens service. To avoid any risk until the implementation of the modification mentioned below, it is furthermore strongly recommended to perform up/down movements of the lifting column not directly above the patient, but complete the vertical movement beside the patient and then move the system horizontally above the patient. We appreciate your understanding and cooperation with this Safety Advisory Notice and ask you to immediately instruct your personnel accordingly. Please ensure that this Safety Advisory Notice is placed in the system's instructions for use until the update has been installed. If you have sold or otherwise disposed of this equipment and it is no under your control, we kindly ask that you forward this Safety Advisory Notice to the new user of the equipment. Please also inform us about the new owner of the equipment. We apologize for any inconvenience this may cause. Further questions please call (610) 219-6300.
- Initiated
- 2014-03-05
- Root cause
- Component design/selection
- Terminated
- 2015-03-16
- Distribution
- US Distribution including MO and OH.
- Recall number
- Z-1460-2014
- Recall status
- Terminated
- CDRH record ID
- 126309
- FDA classified
- 2014-04-16
- Recall event number
- 67780