Skip to content

Public FDA record

Z-1460-2015: EOS, Digital radiography system used in general radiographic examinations.

Recall record Z-1460-2015 for Eos Imaging Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1460-2015
Company in FDA record
Eos Imaging Inc
FDA classification date
May 5, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When performing calibration, an alert message on the spectral filtration of the X-ray beam may be suppressed. Improper filtration of the X-ray Beam can then occur in exams set up with copper filtration.

Source facts

Action
Planned Action of EOS Imaging Inc.: 1. Install a new acquisition software version, which will prevent use of the system when improper filter positioning is detected. 2. A customer notification letter which includes a statement that EOS Imaging will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following CDRH approves the CAP subject to the following conditions: 1. Provide documentation to show that all filter positioning subassemblies have been replaced, which you state was implemented to prevent improper filter positioning. 2. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 3. EOS Imaging will implement this CAP by September 1, 2015. For further questions call (678) 564-5400.
Initiated
2015-02-17
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-05-31
Distribution
US Distribution to the states of: CA, DE, PA, MN, FL, MO, OH, IN and IL.
Recall number
Z-1460-2015
Recall status
Terminated
CDRH record ID
134571
FDA classified
2015-05-05
Recall event number
70732

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗