Open original FDA record ↗Recall record
- FDA identifier
- Z-1460-2017
- Company in FDA record
- Del Mar Reynolds Medical, Ltd.
- FDA record creation date
- Mar 14, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.
Source facts
- Action
- On 08 February 2017, a customer letter will be sent via priority service, return receipt requested, to all U.S. customers. Records of notification will be maintained to ensure complete notification. On 15 February 2017, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Records of notification will be maintained to ensure complete notification. Upgrade is expected to be available in April 2017; all affected Arkon Anesthesia Delivery Systems will receive a software upgrade onsite.
- Initiated
- 2017-02-08
- Root cause
- Software design
- Terminated
- 2017-10-19
- Distribution
- Worldwide Distribution - US nationwide and internationally in the following countries: BOLIVIA, CANADA, CHINA, ECUADOR, FRANCE, IRELAND, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, PANAMA, PERU, POLAND, UNITED KINGDOM.
- Recall number
- Z-1460-2017
- Recall status
- Terminated
- CDRH record ID
- 153199
- FDA record created
- 2017-03-14
- Recall event number
- 76429