Z-1460-2022: Da Vinci X/Xi Large Clip Applier, Model 470230 Da Vinci X/Xi Medium-Large Clip Applier, Model 470327 Da V

Recall record Z-1460-2022 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1460-2022
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Jul 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.

Source facts

Action
On 07/01/2022, Intuitive Surgical sent an "Urgent Medical Device Correction" Letter to customers via FedEx courier informing them there is the potential for unexpected motion while using the da Vinci S/Si and X/Xi Endowrist Clip Applier Instruments. This can happen when one of the da Vinci S/Si and X/Xi Clip Applier grip discs are disengaged from the sterile adapter disc during the engagement routine that occurs prior to insertion of the instrument into the cannula and into the patient. To mitigate this issue, Intuitive Surgical has developed a Clip Applier Manual Self-Test (Appendix A) that is part of the Urgent Medical Device Correction Letter. This test is performed by moving the Clip Applier grips through the full range of yaw motion (side-to-side in the plane of the instrument grips) and observing the movement of both the instrument grips prior to placing a clip around a vessel/tissue. This test should be performed in an open space to prevent any collisions. Pass Criteria " During the yaw movement, if the grips remain open with no observation of grips closing then the instrument passes the self-test and users may proceed with using the clip applier instrument in a procedure. Fail Criteria " During yaw movement, if the grips begin to close as opposed to the grips remaining open then there is potential that one of the grip discs is not engaged. Instrument should be removed, reinstalled, and inspected for engagement. If the self- test fails, remove Clip Applier instrument, re-install and repeat self-test. If it fails again, use a different Clip Applier instrument and return the instrument using the standard Intuitive RMA process. Please call customer service to return the affected product using the appropriate local number listed in section 6 of the letter. Credit will be issued when the reported unexpected motion is confirmed. For questions: " US, Puerto Rico and Dominican Republic: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg
Initiated
2022-07-01
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Recall number
Z-1460-2022
Recall status
Open, Classified
CDRH record ID
194426
FDA record created
2022-07-25
Recall event number
90332

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗