Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
Source facts
Action
A Field Safety Notice was sent to all US consignees via certified mail. Market representatives will distribute the letter internationally. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Initiated
2014-03-27
Root cause
Nonconforming Material/Component
Terminated
2016-02-02
Distribution
Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.
Recall number
Z-1461-2014
Recall status
Terminated
CDRH record ID
126466
FDA classified
2014-04-16
Recall event number
67837