Skip to content

Public FDA record

Z-1461-2020: Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Cat

Recall record Z-1461-2020 for Response Biomedical Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1461-2020
Company in FDA record
Response Biomedical Corp.
FDA record creation date
Mar 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.

Source facts

Action
On 03/15/2019, the firm sent an "Field CORRECTION" Letter via email informing them that the specific lot (4244CK) has seen a decrease in CK-MB levels of up to 57% signal loss within 11 months of the Randox-stated shelf life of 24 months. Customers are instructed to: -Inspect their stock levels and report level of unused inventory; -Notify the Recalling Firm's Technical Support using Field Correction Response Form indicating the replacement quantity; -Confirm lot-specific ranges per website (www.responsebio.com/cardiaccountrols) -Substitute affected Lot # 070318 with replacement stock as appropriate. For any further questions or information contact Technical Support at techsupport@responsebio.com.
Initiated
2019-03-15
Root cause
Device Design
Terminated
2022-06-09
Distribution
Worldwide distribution: US Nationwide distributions in the states of IN, GA, OK, KS, AZ, SC, NE, NC, MI, FL, CA, MN, and AR. Countries of Canada, Italy, Malaysia, Kuwait, Saudi Arabia, UAE, Ukraine, Russia, Dominican Republic, France, Mexico, Colombia, Singapore, China, Serbia, Romania, and Philippine.
Recall number
Z-1461-2020
Recall status
Terminated
CDRH record ID
178445
FDA record created
2020-03-10
Recall event number
84514

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗