Open original FDA record ↗Recall record
- FDA identifier
- Z-1462-06
- Company in FDA record
- Becton Dickinson & Company
- FDA classification date
- Sep 9, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm received complaints indicating failure of the Non-Patient (NP) sleeve to function properly. This sleeve covers the cannula and prevents leakage during blood collection.
Source facts
- Action
- BD sent out recall notifications on 6/21/2006 to distributors and their direct account customers.
- Initiated
- 2006-06-21
- Root cause
- Other
- Terminated
- 2007-06-07
- Distribution
- World wide-Products are distributed nationwide to distributors and direct accounts. The products are also distributed internationally to the following locations: Japan, Australia, Canada, Hong Kong, New Zealand. For European distribution, the products are shipped to Temse located in Belgium who handles the recall. Temse has been identified as recalling only the following products and lots: Catalog Number 364896 (Direct Draw Adapter), lots 5311376, 5312193, 5312201. Catalog Number 367300 (Luer Adapter), lots 5307894, 5307896, 5307900. Government accounts include the VAMC in Erie, PA;
- Recall number
- Z-1462-06
- Recall status
- Terminated
- CDRH record ID
- 46976
- FDA classified
- 2006-09-09
- Recall event number
- 35831