When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Source facts
Action
The firm sent a notice regarding a Critical Bug in nordicBrainEx version 2.2.1 to 2.3.10 on 06/16/2022 by email. The notice explained the problem and advised users to use extra caution when viewing the exported results. A software revision will be released to correct the issue.
Initiated
2022-06-16
Root cause
Software design
Terminated
2024-09-04
Distribution
US Nationwide distribution.
Recall number
Z-1462-2022
Recall status
Terminated
CDRH record ID
194531
FDA record created
2022-07-25
Recall event number
90480