Skip to content

Public FDA record

Z-1462-2023: MAQUET CARDIOSAVE hybrid Intra-Aortic Balloon Pump, Model Number 0998-XX-0800-32

Recall record Z-1462-2023 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1462-2023
Company in FDA record
Datascope Corp.
FDA classification date
Apr 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.

Source facts

Action
An URGENT FIELD SAFETY NOTICE MEDICAL DEVICE CORRECTION was sent to customers beginning 4/4/23. Actions to be taken by the Customer now: Our records indicate that that you have received one or more CARDIOSAVE hybrid IABPs having the serial numbers that are affected by this recall. 1. Inspect the IABP to determine if it has a Type J power plug (refer to Image 1 below). If present, the IABP should not be used until the cord reel has been replaced. 2. Schedule a service appointment with your Datascope Corp./Getinge Service Representative for the cord reel correction from a Type J connector to Type N (refer to Image 2 below). 3. Please ensure that all Cardiosave Intra-Aortic Balloon Pump users at your hospital / facility are aware of this notice. 4. Please forward this information to all current and potential Cardiosave Hybrid IABP users within your hospital/facility. 5. If you are a distributor who has shipped any affected products to customers, please forward this letter to their attention for appropriate action. Actions to be taken: Please examine your inventory to determine you have a Cardiosave Hybrid IABP. Whether you have an affected cord or not, please complete and sign the attached RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Actions to be taken by Getinge: Getinge will update the affected Cardiosave Hybrid IABPs that contain a Type J connector with a Type N connector to correct and address this issue. This voluntary correction notification only affects the products listed on page 1; no other products are affected by this voluntary correction. We apologize for any inconvenience this Medical Device Correction may cause. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope Corp./Getinge Representative or office.
Initiated
2023-04-04
Root cause
Component design/selection
Distribution
International distribution in the country of Brazil.
Recall number
Z-1462-2023
Recall status
Open, Classified
CDRH record ID
199426
FDA classified
2023-04-26
Recall event number
92021

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗