Open original FDA record ↗Recall record
- FDA identifier
- Z-1462-2024
- Company in FDA record
- Agfa N.V.
- FDA classification date
- Apr 2, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury
Source facts
- Action
- AGFA issued Urgent Field Safety Notice via email on 2/23/24. Letter states reason for recall, heallh risk and action to take: Agfa/local business partner will inspect the cable of your device at the earliest opportunity and will replace it in case of any potential malfunction of the safety system. If you have any questions about this matter, please contact your local Agfa NV organization/local business partner.
- Initiated
- 2024-02-23
- Root cause
- Component design/selection
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-1462-2024
- Recall status
- Open, Classified
- CDRH record ID
- 206428
- FDA classified
- 2024-04-02
- Recall event number
- 94143