Skip to content

Public FDA record

Z-1462-2024: Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ra

Recall record Z-1462-2024 for Agfa N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1462-2024
Company in FDA record
Agfa N.V.
FDA classification date
Apr 2, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury

Source facts

Action
AGFA issued Urgent Field Safety Notice via email on 2/23/24. Letter states reason for recall, heallh risk and action to take: Agfa/local business partner will inspect the cable of your device at the earliest opportunity and will replace it in case of any potential malfunction of the safety system. If you have any questions about this matter, please contact your local Agfa NV organization/local business partner.
Initiated
2024-02-23
Root cause
Component design/selection
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Kingdom.
Recall number
Z-1462-2024
Recall status
Open, Classified
CDRH record ID
206428
FDA classified
2024-04-02
Recall event number
94143

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗