Z-1463-2017: The ABL800 Model#:All

Recall record Z-1463-2017 for Radiometer America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1463-2017
Company in FDA record
Radiometer America Inc
FDA record creation date
Mar 14, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports that the sample type "cord blood" has been changed into the two sample types "cord blood arterial" and "cord blood venous.

Source facts

Action
Customer notification letters were sent 01/27/16. The letter provided the following instructions: The firm recommends that customers ensure that operators and clinicians using the "cord blood sample results from a HIS/LIS or a middleware system to determine treatment are made aware that arterial or venous is not transmitted. An upgraded version of the ABL90 and ABL800 series software will ensure that the full sample type cord blood arterial and cord blood venous will be transmitted to the HIS/LIS or middleware system. The new software version will be installed by local engineer when available. If not the end user customers are instruction to provide a copy of the recall notification letter to the end user. Any questions regarding this letter contact Radiometer America Technical Support at 1-800-736-0600 opt.1.
Initiated
2017-01-26
Root cause
Device Design
Terminated
2018-02-20
Distribution
U.S. distribution to the following; nationwide. Foreign distribution to the following; Netherlands, China, Czech Republic, Denmark, Germany, Spain, France, Hungary, India, Japan, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland.
Recall number
Z-1463-2017
Recall status
Terminated
CDRH record ID
152860
FDA record created
2017-03-14
Recall event number
76339

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗