Open original FDA record ↗Recall record
- FDA identifier
- Z-1463-2026
- Company in FDA record
- Medline Industries, LP
- FDA classification date
- Feb 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Source facts
- Action
- On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.
- Initiated
- 2026-01-08
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1463-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218311
- FDA classified
- 2026-02-24
- Recall event number
- 98333