A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
Source facts
Action
.The firm, Phillips, sent an " URGENT Medical Device Correction" letter dated
2013 March 06, to its costumers . The letter disc ribes the product, problem and actions to be taken. The costumers were instructed to Verify that all planned image reconstructions are completed, and Perform off-line reconstructions if expected images are missing. The firm will release A Field Change Order (FCO) to correct the issue with software version release 3.5.7.and A Philips Field Service Engineer will contact you to schedule the software installation at your site.
If you need any further information or support concerning this issue, please
contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Initiated
2014-03-06
Root cause
Software in the Use Environment
Terminated
2014-11-05
Distribution
Worldwide Distribution:US ( Nationwide) Including stages of MI, OR, PA, TX & WV; and Countries of: Australia, Bangladesh, Belgium, Brazil, Bulgaria, China, France, Germany, India, Indonesia, Ireland, Japan, Nicaragua, Oman, Palestine, Russia, South Korea, Spain, Switzerland, Thailand, Turkey & United Kingdom.
Recall number
Z-1464-2014
Recall status
Terminated
CDRH record ID
126256
FDA classified
2014-04-16
Recall event number
67763