Open original FDA record ↗Recall record
- FDA identifier
- Z-1464-2017
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Feb 28, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter because of reports of separation and/or leakages between the Y-connector and tubing.
Source facts
- Action
- An Urgent Medical Device Recall Notification letter will be sent to customers on February 20, 2017 to inform them that CareFusion is recalling the MaxGuard Extension Set with Injection Site and 0.2 micron filter model number ME3305 Lot Number 16016790. Customers are informed that CareFusion has received reports of separation and/or leakages between theY-connector and tubing. Customers are informed that leakages from an extension set can cause delay of infusion, interruption of infusion, exposure to medication or hazardous infusates, or underinfusion. The letter informs the customers of the actions to be taken and for recall related questions to contact CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact customer advocacy at (888) 812-3266.
- Initiated
- 2017-02-20
- Root cause
- Process control
- Terminated
- 2017-11-17
- Distribution
- US Distribution including MA and MD
- Recall number
- Z-1464-2017
- Recall status
- Terminated
- CDRH record ID
- 153214
- FDA classified
- 2017-02-28
- Recall event number
- 76439