Skip to content

Public FDA record

Z-1464-2022: AquaC UNO H [120V, 60Hz, 1000VA, Fresenius US-Version: 24-10VA-1 Vivonic Part Number: G02000970-US

Recall record Z-1464-2022 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1464-2022
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Jul 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal ethernet cable may be misaligned, the cable jacket (insulation) could become compromised due to high temperature and result in the internal ethernet cable conducting electrical current to the ethernet port on the outside of the device), potentially harming service technicians and/or other authorized operators. If the ethernet port on the device is touched or manipulated, there is a risk of electrical shock

Source facts

Action
Fresenius Medical issued Urgent Medical Device Correction letter (FA-2022-16-C) on 6/16/22. Letter states reason for recall, health risk and action to take: DO NOT TOUCH THE ETHERNET PORT ON THE BACK OF THE MACHINE. " If your ethernet port cover is not covered by the enclosed cover, unplug the machine, and carefully cover it without touching the ethernet port. " Do not perform any service to the device requiring you to connect an ethernet cable to the device. Do not attempt to download FTP files via the ethernet cable. " You may continue to use the device until the device correction is made. A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. If choosing to correct via your own Certified AquaC UNO H System Technician, you will need to contact FMCRTG prior to performing the correction. FMCRTG will provide the necessary instruction via a Change Notification. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
Initiated
2022-06-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1464-2022
Recall status
Open, Classified
CDRH record ID
194643
FDA classified
2022-07-25
Recall event number
90527

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗