Skip to content

Public FDA record

Z-1464-2023: VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS

Recall record Z-1464-2023 for Covidien LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1464-2023
Company in FDA record
Covidien LP
FDA classification date
Apr 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated March 2023 was sent to customers. Customer Actions: Our records show that your facility has received the impacted product. Medtronic requests that you take the following actions: " Identify and quarantine any used and unused impacted product. Please refer to Attachment A: Impacted Serial Number Lists for a list of affected products. For verifying the respective product and serial number on the device and packaging please refer to Attachment B. " Sterilize all used devices as per your standard decontamination protocol prior to returning used cannulas. Utilization of appropriate personal protective equipment is advisable during decontamination of the device. " Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. " Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. " Forward this notification to any entities within your organization or to any other organization where the potentially affected product has been transferred or further distributed. Adverse events or quality problems experienced with this product should be reported to the FDA and Medtronic: " Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm, or call FDA at 1-800-FDA-1088 (1-800-332-1088) " Medtronic Technical Services at 1-800-962-9888, option 2 to provide information regarding those events. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Field Representative.
Initiated
2023-03-22
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of NC and the countries of Belgium, Denmark, France, Germany, India, Italy, Netherlands, Spain, Sweden.
Recall number
Z-1464-2023
Recall status
Open, Classified
CDRH record ID
199399
FDA classified
2023-04-26
Recall event number
92004

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗