Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
Source facts
Action
BioMerieux issued an URGENT FIELD SAFETY NOTICE to its consignees on 03/14/2024 via email. The notice explained the issue, impact to patient, and provided the consignee (user) with alternate testing methods.
BioMerieux issued another URGENT FIELD SAFETY NOTICE dated 07MAY2024 to its consignees on May 9, 2024, via email. This letter includes the statement, "however, no results for the Ceftriaxone (cro02n) should be reported outside the lab."
Initiated
2024-03-14
Root cause
Employee error
Distribution
Worldwide
Recall number
Z-1464-2024
Recall status
Open, Classified
CDRH record ID
206759
FDA classified
2024-04-25
Recall event number
94264