Open original FDA record ↗Recall record
- FDA identifier
- Z-1465-2017
- Company in FDA record
- Smiths Medical ASD Inc.
- FDA record creation date
- Mar 14, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Smiths Medical became aware that one lot of the Bivona 3.5mm TTS Cuffed Pediatric tracheostomy tubes was incorrectly labeled for neonatal use instead of correctly labeled for pediatric use.
Source facts
- Action
- Consignees were sent a Smiths Medical "Urgent Medical Device Field Safety Notice" dated February 08, 2017. The letter was addressed to "Clinician". The letter listed the Affected devices, Reason for Recall, Risk to Health, and Instructions to Customers. Advised customers to complete and return the "Urgent Medical Device Field Safety Notice (FSN) Response Form" to smithsmedical4390@stericycle.com. All affected devices must be returned to Stericycle for processing. If the product was further distributed, consignees were asked to immediately notify them of the Recall and retrieve all affected devices. For questions contact Stericycle via e-mail at the following address: smithsmedical4390@stericycle.com.
- Initiated
- 2017-02-09
- Root cause
- Labeling False and Misleading
- Terminated
- 2018-06-20
- Distribution
- US: CA, ILM NC, GA, NE, MA, CO, CT, MO, TN, LA, MD, MN, FL, SC,M OK, MI, PA, TX, WA, WI.
- Recall number
- Z-1465-2017
- Recall status
- Terminated
- CDRH record ID
- 153205
- FDA record created
- 2017-03-14
- Recall event number
- 76433