Z-1466-06: FP1000 Cell Preparation System Part Number 624922

Recall record Z-1466-06 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1466-06
Company in FDA record
Beckman Coulter Inc
FDA classification date
Aug 31, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Cell Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results.

Source facts

Action
A Product Corrective Action letter was sent on June 26, 2006 informing users that during the cleaning cycle performed during the shutdown procedure of the Beckman Coulter FP1000 Preparation System, fluid (diluted bleach) may drip from the probe in the location of the back reagent rack potentially resulting in bleach and/or water dripping into the reagents contaminating them and resulting in possible incorrect results. To reduce the risk of incorrect results from dripping fluid, users were advised to remove all patient samples, reagents, controls, and calibrators before initiating the shutdown cycle of the Beckman Coulter FP1000. Software will be modified in the next version to prevent the possibility of dripping bleach during the shutdown cycle. The Product Corrective Action letter is to be maintained as part of the customer''s laboratory Quality System documentation. The enclosed Response Form is to be completed and returned within 10 days to the recalling firm. Customers are also advised to contact the firm or their local Beckman Coulter Representative if there are questions.
Initiated
2006-06-26
Root cause
Other
Terminated
2012-04-26
Distribution
Nationwide and Canada
Recall number
Z-1466-06
Recall status
Terminated
CDRH record ID
46858
FDA classified
2006-08-31
Recall event number
35778

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗